5-12
Operation Manual
3 Points Angle
This is a basic measurement that is available in all diagnosis modes. Place three points on a 2D image
and measure the angle formed by the points.
1. Tap 3 Points Angle on the touch screen.
2. Specify three points.
3. The angle formed by the three points will be calculated and displayed on the screen.
2 Lines Angle
A basic measurement that is available in all diagnosis modes. Specify two straight lines in a 2D image
and measure the angle between them.
1. Tap the 2 Lines Angle button on the touch screen.
2. Draw two straight lines. For information on drawing a straight line, please refer to ‘Distance.
3. The angle between two lines will be calculated and displayed on the screen.
X
When two angles are calculated, the smaller angle is displayed.
NOTE: You can select and measure 3 Points Angle or 2 Lines Angle.
%StD
StD stands for Stenosis Distance, which is a basic measurement available in all diagnosis modes. In a
2D image, the diameter of a vessel is measured and the stenosis ratio calculated.
1. Tap the %StD button on the touch screen. “%StD” is displayed in the user information area.
2. Measure the total diameter of a vessel using the Distance measurement method.
3. When a new cursor appears, measure the inner wall diameter of the vessel under stenosis.
4. Calculate %StD with the following equation:
%StD = (Outer Distance – Inner Distance) / Outer Distance x 100