Operation Manual
Simultaneous
Tap Simultaneous on the touch screen. Each time you tap the button, the Simultaneous function
turns on and off.
NOTE: It appears in the PW menu only when “Utility > Setup > General > Simultaneous Mode” is
set to Allow.
If the Simultaneous function is enabled, you can view the 2D image and Spectral Doppler image
simultaneously in real time. If it is disabled, you can view only one of the two images.
The Simultaneous function decreases Doppler PRF, thus decreasing the measurable speed range.
Invert
Tap Invert on the touch screen.
Each time the button is pressed, the speed indicator (+ / –) for a spectrum is inverted.
Sound
Adjusts the doppler volume. Select a value between 0 and 100 by tapping the button or rotating
the dial-button.
Mean Trace
Tap Mean Trace on the touch screen. Mean Trace for the selected spectrum is performed.
Auto Clac
Turn it on or off by pressing Auto Clac on the touch screen.
If it is turned on, Doppler Trace is performed and its results are displayed. . The values selected from
Setup > AutoCalc are displayed on the screen. You can select up to six values. If it is turned on, the
Mean Trace button is enabled.
For information on setting display values, please refer to Chapter 7, “Utility.
CAUTION: The measurements done by Auto Trace under Measure and Real Time Automatic
Doppler Trace (Automatic Calculator) may be different from each other. This is because the
algorithms for these two methods are different. It is recommended to use Auto Trace under
Measure for more accurate measurement.