4-10
Operation Manual
■
Panoramic (Optional)
NOTE: The Panoramic Imaging function is available only in 2D mode with Linear and Convex
probes.
Tap Panoramic on the touch screen. The touch screen will be switched to the Panoramic Ready
screen. For more information, please refer to “Panoramic” in this chapter.
■
Low MI
NOTE: Low MI is available under 2D mode with specific probes only.
Turn it on or off by tapping Low MI on the touch screen. The monitor changes to Low MI screen.
Refer to ‘LOW MI Mode’ in this chapter for detail.
■
Preset
Tapping Preset on the touch screen displays the Probe settings supported by the present
application. For more information, please refer to Chapter 7 “Utilities”.
■
Spatial Compound (Optional)
NOTE: This item appears in the Flexible Soft menu only when a Linear Probe is used.
Tap Spatial Comp on the touch screen. The Spatial Compound mark appears on the image
information. Spatial Comp, which can be used to select a value for Spatial Compound, is enabled
on the touch screen. Tap the button on the touch screen or use the dial-button to select Low,
Medium, or High for Spatial Compound.
■
SRF (Speckle Reduction Filter)
Turn it on or off by tapping SRF on the touch screen. ‘SRF’ will be displayed in the image data.
SRF stands for Speckle Reduction Filter. It is used to optimize an image by minimizing noise and
adjusting the brightness of boundaries automatically. If it is turned on, SRF Index is enabled on the
touch screen.