Vista de pagina 131
4-24
Operation Manual
C Mode Menu
Invert
Tap Invert on the touch screen. The color bar is inverted each time this button is tapped. When the
color bar is inverted, the colors on the image are also inverted.
Frequency
Set the probe frequency by tapping Frequency on the touch screen or using the dial-button. Select
Pen or Gen by tapping the button.
X
Gen (General): General frequency
X
Pen (Penetration): Low frequency
Since the selected frequency is displayed in the title area, you can easily see the current state of the
frequency.
Filter
Select a value between 0 and 3 by tapping Filter on the touch screen or using the dial-button.
This is an electric filter to remove the low-frequency doppler signals generated by the movement
of blood vessel walls. Adjust the Cutoff Frequency to remove doppler signals whose frequency is
lower than the cutoff frequency.
Scale (PRF)
Rotate the Scale dial-button on the touch screen. Rotating the Scale dial clockwise makes the PRF
(Pulse Repetition Frequency) increase so that the speed range of the blood flow is widened, and
vice versa.
BaseLine
Rotating the dial-button clockwise increases the Baseline.
In Color Doppler mode, the color bar indicates the direction and speed of blood flow. Based on the
Baseline at the middle, red indicates the direction and speed of blood flow toward the probe. By
contrast, blue indicates the direction and speed of blood flow that flows in the opposite direction
of the probe.