4-9
Chapter 4 Diagnosis Modes
NOTE: The Harmonic function is available with specific probes only.
Trapezoidal
NOTE: The Trapezoidal item appears in the menu only when a Linear Probe is used.
Turn it on or off by tapping Trapezoidal on the touch screen.
In general, the rectangular frame provided by a Linear Probe is changed to a trapezoidal shape. This
allows a wider view of an image.
The Trapezoid function may not be available for certain depths. In addition, the Write Zoom function
cannot be accessed with the Zoom button when the Trapezoidal function is in use. (However, Read
Zoom can be used.)
2D/C Live
Turn it on or off by tapping 2D/C Live on the touch screen.
The 2D image and Color Doppler image for the scanned area can be displayed simultaneously in
real time.
In ACCUVIX XG this item is placed in the 2D menu so that the 2D/C Live Mode, which is used
frequently, can be easily accessed. You can toggle between two images easily with a single button
operation.
Pulse Inversion
Turn it on or off by tapping Pulse Inv on the touch screen.
If it is turned on, pulses are inverted to sharpen the displayed image. The INV mark appears at the
top of the image.
NOTE: Pulse Inversion is available with specific probes only.