4-46
Operation Manual
■
Contrast
Select the contrast for an E image between 1 – 100%.
■
Enhancement
Control brightness and sensitivity of E image between 1 – 100%. A higher value provides more
accurate images of boundaries. However noise may increase according to the value.
■
Color Map Index
Select the color for an E image between Type 1 - Type 5. Changing the color map also change the
color bar.
■
Blending Level
Select the ratio of Alpha Blending. You can adjust the ratio in 5% steps between 0 - 100%. Setting it
to 0% shows an E image only and setting it to 100% shows a 2D image only.
■
Persistence Level
Specify the change speed of frame between frames. You can adjust the ratio in 5% steps between
0 – 100%. If you set a higher value, the image applies the previous frame much more.
NOTE:
X
The Following are not available or have limit to use in E mode;
−
Not available functions: Angle, Scan Area, ECG, Histogram.
−
Available only one Focus.
−
Available Biopsy On and Off only.
−
2D measurements, including measurement of distance, circumference and area, are available
only.
X
If you press the Dual or Quad button in E mode, it will terminate E mode and enter to 2D
mode..