4-88
Operation Manual
Slice Thick.
NOTE: This is used only when Cut Type is Plumb.
Adjust the interval of the vertical lines by tapping the button on the touch screen or using the dial-
button. You can adjust in 1 mm units from the center line, and the current interval is displayed in the
image information.
Rotate Line
NOTE: This cannot be used when:
X
Cut Type is Contour or
X
Cut Type is Line and Selected Slice is Select All.
Rotate a straight line by tapping the button on the touch screen or using the dial-button. When the
line is rotated, the oblique image is also changed accordingly.
OVIX
NOTE: This cannot be used when Cut Type is Contour.
OVIX is the abbreviation for Oblique View eXtended which sets the cross-sectional thickness of an
oblique image and shows the image in 3D.
Tap this button on the touch screen to switch this function on or off. When this is on, an OVIX Line
appears in the reference image which indicates the cross-sectional thickness of the oblique image of
the reference image. OVIX Post Curve is also displayed at the left of the monitor screen.
The thickness of the OVIX Line can be adjusted by OVIX Thick. To change the 3D image settings, select
and adjust Render Setup or OVIX Post Curve from the 3D Utility Menu.