Operation Manual
HPRF (High PRF) Function
Adjust the blood flow above the speed limit at a specific depth. This increases the scale. This function
is available only in PW Spectral Doppler Mode (D Only).
X
Activating HPRF
To activate HPRF, increase the Scale values at the required depth. A The Phantom Gate will
appear on the D Line at a position higher than the sample volume. Once HPRF starts, PRF does
not increase even if you increase the scale value.
X
Finishing HPRF
While HPRF is in use, decrease the scale value by one step to finish HPRF. The maximum PRF
values in PW Spectral Doppler Mode are shown.
X
Moving Sample Volume
To move the Sample Volume position in the D Only state, the system calculates PRF values and
the Phantom Gate position, and updates them on the PW Spectral Doppler image. HPRF is
terminated when HPRF cannot be activated.
When Sample Volume is moved in the 2D Only state, the PRF values don’t change.
NOTE: HPRF is not activated in Simultaneous Mode. This is set in [Utility] > [General] > [Scan
Mode] > [Simultaneous]. It is also not activated if [PRF*2] exceeds 23KHz.
CAUTION:
X
The Phantom Gate position can be located outside the 2D image area in Zoom Mode.
X
Make sure that sample volume and Phantom Gate are not placed together in the measuring
area. If more than two Sample Volumes are located in the vessels, all Doppler components will
appear in the spectrum, causing noise.