4-99
Chapter 4 Diagnosis Modes
Acquiring XI STIC Image
NOTE:
X
When a heart rate cannot be calculated because motion artifact is severe or a cardiac cycle is
not present, the system returns to the initial XI STIC screen.
X
In Color STIC, the cardiac cycle can be measured only when the frame rate is 20 or above.
1. Select the XI STIC tab in the 3D Stand By screen on the touch screen.
2. Set various parameters as you would for acquisition of standard 3D images.
3. Press the Freeze or Set button on the control panel. The system will start acquiring 3D images.
4. When image acquisition is complete, XI STIC is displayed on the monitor screen and the XI STIC
Confirm screen appears on the touch screen. Check the fetal cardiac cycle calculated.
5. Tap Yes on the touch screen to continue. Tap No to cancel and acquire images again.
6. Optimize and diagnose the acquired images.
Tips!
How to Improve STIC Volume Data
X
Scan Angle: Specify a small scan angle for small fetal hearts.
X
Scan Position: Adjust the scan position so that the center of the scan angle and the fetal heart
are aligned properly.
X
ROI Box: Adjust the size of the volume box so that it nearly fits the size of the fetal heart.
Scan Angle
Set the scan angle. Select between 15 and 60˚by clicking the corresponding button or using the dial-
button on the touch screen.
Scan Time
Set the image acquisition time. Select a value between 7s and 15s by clicking the corresponding
button or using the dial-button on the touch screen.