4-16
Operation Manual
X
Chroma Map: Select from Type 1 to Type 13 or from User 1 to User 3. You can change the settings
for the user types in Utility > Post Curve > 2D Post > Pseudo Color. For more information, refer
to ‘Post Curve’ in Chapter 3. “Settings”.
X
Gray Map: Select from Type 1 to Type 13.
X
Dynamic Range: Select a value between 50 and 170. This is used to adjust the contrast of the
image by changing the minimum and maximum ratios for input signals.
X
Reject Level: Select a value between 1 and 32. This is used to eliminate noise and echoes from
the image.
[Figure 4.3 Low MI - Touch Screen]
Panoramic (Optional)
NOTE: The Panoramic function is available only in 2D mode with Linear and Convex probes.
Panoramic Imaging is the function that acquires images over a wider range by using continuous
ultrasonographic images. Up to 500 frames can be used.
Acquiring a Panoramic Image
1. Tap Panoramic on the touch screen. The touch screen will be switched to the Panoramic Ready
screen.
2. Tap Start/Stop on the touch screen. The system will begin to acquire a panoramic image.