GE HEALTHCARE
DIRECTION 5194296-100, REVISION A LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
10-6 Section 10-5 - System Maintenance
10-5-2 Functional Checks (See Also Chapter 4)
The functional checks take about 60 minutes to perform. Refer to the system user documentation
whenever necessary.
10-5-2-1 System Checks
NOTE: * Some software may be considered standard depending upon system model configuration.
Table 10-5 System Functional Checks
÷ Step Description
B-Mode
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode of
operation.
CF-Mode*
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that affect this
mode of operation.
Doppler Modes*
Verify basic Doppler operation (PW if available). Check the basic system controls that affect this mode
of operation.
M-Mode
Verify basic M-Mode operation. Check the basic system controls that affect this mode of operation.
*Applicable
Software Options
Verify the basic operation of all optional modes such as Multi-Image, 3D*, Harmonics, Cine,... etc.
Check the basic system controls that affect each options operation.
Xmit/Recv
Elements
Use the Visual Channel Utility on the loop connect to verify that all system xmit/recv channels are
functional.
Keyboard Test
Perform the Keyboard Test Procedure to verify that all keyboard controls are OK.
LCD
Verify basic LCD display functions. Refer to Chapter 3 of the User Manual.
Software Menu
check
Verify Software Menu display functions. Refer to Chapter 3 of the User Manual.
Measurements
Scan a gray scale phantom and use the measurement controls to verify distance and area calculation
accuracy. Refer to the User Manual, Chapter 18, for measurement accuracy specifications.
* 3D is option on LOGIQ Book XP Vet, but not available on LOGIQ Book XP PRO.
* CF and Doppler mode are option on LOGIQ Book XP PRO/XP Vet.