GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
8-2 Section 8-2 - Disassembly/Re-assembly
8-2-2 Handle Assy (FRU No. 312)
Purpose: This is a description on how to remove and replace the Handle Assy (FRU No.312).
8-2-2-1 Tools
• Plier
8-2-2-2 Needed Manpower
• 1person, 2 minutes + travel
8-2-2-3 Preparations
• Shut Down the System, disconnect the AC/DC power cord and remove the battery.
8-2-2-4 Removal Procedure
Refer to Figure 8-1 on page 8-3 .
1.) Unscrew the two handle caps on both sides of the system, the rotation direction is
counterclockwise.
2.) Pull out the Handle.
8-2-2-5 Mounting procedure
1.) Install the new parts in the reverse order of removal.