GE HEALTHCARE
D
IRECTION 5194296-100, REVISION A LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 5 Components and Functions (Theory) 5-5
5-3-3 Air Flow Distribution
The three air flow passes allow the system to be cooled down as shown in the figure above.
Pass A (Rear Fan> MST, DFP Assy> Bottom front/Bottom left) for MST, DFP Assy cooling.
Pass B (Bottom> Bottom Fan> DCDC> Bottom left) for DCDC cooling.
Pass C (Bottom> CPU Fan> Bottom left) for CPU cooling.
Figure 5-3 Air Flow Inside the System
A
A
B
B
C
C