GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
3-44 Section 3-9 - Loading the System Software
3-9-5 Option Key Update
NOTE: After the system software loading completion, please input option key and check the
option strings to ensure that the options are activated and working.
NOTE: Option key is required to be input when the system is powered on after software loading
completion.
1.) Power on the system.
2.) Input series number of the system and select “OK“.
3.) Input option key according to SW Option Key label.
4.) Press “OK” and the system enters scanning mode.
5.) Refer to the process as below to verify if the options are activated and working.
Select Utility -> Advanced -> Admin -> System Admin , then check “option” list. If the option is
enable or permanent, it means the option function is available; if the option is disabled, it means the
function is invalid.
Figure 3-44 Option key screen