Samsung SV-651B Manual de servicio Pagina 67

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 226
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 66
GE HEALTHCARE
D
IRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 3 Installation 3-11
3-5-2-2 System Shutdown
Disconnect the Mains Power Cable if necessary. For example: Relocating the system.
3-5-3 Transducer Connection
1.) Carefully open the system LCD display, plug the probe connector into the probe port, then lock the
probe latch upward.
NOTE: Please ensure that the probe latch is in an unlocked position before you connect the probe to the
systsem.
Figure 3-8 Connect the probe
NOTE: It is not necessary to turn OFF power to connect or disconnect a probe.
probe latchprobe latch
Vista de pagina 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 225 226

Comentarios a estos manuales

Sin comentarios