Operation Manual
1-8
CAUTION:
X
The system has been designed for 100-120VAC and 200-240VAC; you should select the input
voltage of printer and VCR. Prior to connecting a peripheral power cord, verify that the voltage
indicated on the power cord matches the voltage rating of the peripheral device.
X
An isolation transformer protects the system from power surges. The isolation transformer
continues to operate when the system is in standby.
X
Do not immerse the cable in liquids. Cables are not waterproof.
X
The auxiliary socket outlets installed on this system are rated 100-120V and 200-240V with
maximum total load of 200W. Use these outlets only for supplying power to equipment that is
intended to be part of the ultrasound system. Do not connect additional multiple-socket outlets
or extension cords to the system.
X
Do not connect any peripheral devices not listed in this manual to the auxiliary socket outlets of
the system.
X
Do not touch SIP/SOP and the patient simultaneously. There is a risk of electric shock from
leakage current.