Operation Manual
7-6
Display
To set the information about images and related data, select the Display tab in the System menu.
Auto Freeze
After the preset time span (Minute) of inactivity, the scan mode is automatically frozen.
Screen Saver
After the preset time span (Minute) of inactivity, the screen saver is automatically started.
Post Map
This sets the display of the Post Map in the Feedback section at the bottom of the screen.
TGC Line
This sets whether or not the TGC line is displayed. If ‘O is selected, the TGC Line is not shown. If ‘O
after 3 seconds’ is selected, the TGC value appears when a TGC value is adjusted, but disappears after 3
seconds. If ‘On is selected, the TGC Line is always shown.
TI (Thermal Index)Display
The system sets TI values automatically. However, this menu allows the user to choose manually from
one of the three TI parameters as desired: Default, TIs or TIb.
HPRF
Enable or disable High Pulse Repetition Frequency (HPRF) supported in the PW Spectral Doppler mode.
If it is set to ‘On, HPRF is supported by default.
Bodymarker After Freeze
Determine whether the system will automatically switch to the Body Marker mode when the Freeze
button is pressed. If it is set to ‘On, a Body Marker appears when the Freeze button is pressed during
scanning. If it is set to ‘O, a Body Marker appear only when BodyMarker is pressed during scanning.
2D Image Size
The Image size can be set to 100%, 95% or 90%.