Operation Manual
4-16
Use the Change button to adjust the position and size of the ROI box. Each time you press the Change
button, the status of the ROI box is shown in the lower left corner of the screen as follows:
C Pos.: You can move the ROI box. Use the trackball to move the ROI box and press the Set button to
conrm its new position.
C Size: You can resize the ROI box. Use the trackball to resize the ROI box, press the Set button to
conrm its new size.
Color Bar
In Color Doppler Mode, the color bar indicates the direction and speed of blood ow. Relative to the
baseline in the middle, the red color indicates the direction and speed of blood flowing toward the
probe. Conversely, the blue color indicates the direction and speed of blood flowing away from the
probe.
C Mode Menu
Following items are displayed on menu when you press the Menu dial-button in C mode. Select the item
and adjust the value with the Menu dial-button.
Balance
Adjusts the range of the color image displayed by comparing the gray level of the 2D image with the
Doppler signal value of the color image. If the Balance value increases, the color image is seen even
where the 2D image signal is strong (bright), thereby increasing the color image range.
Select a value between 1 - 16 with the Menu dial-button.
Sensitivity
Set the sensitivity of the color image. Increasing the value enhances the color image, but reduces the
frame rate. Select a value between 8 - 31 with the Menu dial-button.
Invert
Inverts the color bar in the opposite direction. Inverting the color bar also inverts the colors displayed
on the image.