5-28
Operation Manual
Volume Flow Measurement
Select Volume Flow in the measurement menu.
Volume Flow allows you to measure and calculate an area or distance. For information on distance or
area measurements, please refer to ‘Basic Measurements. The TAMV (Time Avg. Mean Velocity) value is
automatically measured.
Vesl. Area (Vessel Area)
Measure the area of a blood vessel and calculate TAMV and Volume Flow.
Vesl. Dist.(Vessel Distance)
Measure the width of a blood vessel and calculate TAMV and Volume Flow.
Stenosis Measurement
You can measure the stenosis of each blood vessel system by measuring and calculating an area or
distance.
% StA
Measure the area of the inner and outer walls of a blood vessel. StA stands for Stenosis Area.
1. Select the %StA menu and the rst cursor will appear in 2D Mode.
2. Measure the area of the vessel’s outer wall using the Circ/Area measurement method.
3. When the second cursor appears, measure the area of the inner wall of the vessel that is under
stenosis.
%StA = (Outer Area – Inner Area) / Outer Area x 100