4-15
Chapter 4 Diagnosis Modes
Negative
Inverts the M image’s colors. Press the soft menu’s dial-button [2] to turn this function on or o.
Loop Size
Used to adjust the M image’s size. Set to a value between 30 and 70 by rotating the [3] dial button
beneath the soft menu.
Chroma Map
Press the soft menu’s dial-button [4] to turn this function on or o. Once turned on, rotate the soft
menu’s dial-button [4] and set the image color to Type 1 through 13, or User 1 through 3.
NOTE: User types can be changed from Utility > Post Curve > M Post > Chroma Map. For more
information, refer to ‘Chapter 7. Utilities,’ and specifically the ‘Post Curve’ section.
M Edge Enhance
Used to select the M image’s edge enhancement value. Set to a value between -3 and 3 by
rotating the [6] dial button beneath the soft menu. Higher values provide more accurate images
of boundaries.
Anatomical M
NOTE: Available only with the phased array probe or the cardiac application.
Press the soft menu’s dial-button [6] to turn Anatomical M on or o. Once turned on, you can
change the M line length. Use the Change button to reposition M Point 1 and M Point 2.
NOTE: For more information on other soft menu options and functions, refer to the 2D Mode
section.
Comentarios a estos manuales