5-14
Operation Manual
D A/B
This is a basic measurement that is only available in Spectral Doppler Mode. You can specify two
points in a Spectral Doppler image and measure the velocity at each point to calculate the ratio of the
velocity between them.
1. Press Caliper and select D A/B under soft menu [1]. ‘D A/B is displayed in the user information area.
2. Specify two points for which to measure velocity.
X
Use the trackball to place the cursor at the desired position and press the Set button.
3. When the measurement is nished, its result is shown on the screen.
X
V1: Velocity at point 1
X
V2: Velocity at point 2
X
PGmax: Max Pressure Gradient
X
V1/V2: Velocity ratio
NOTE: If Application under Utility > Measure Setup > General > Caliper is set to Cardiac, results
for Vmax, PGmax, V2-V1, etc. are shown on the screen.
D Trace
This is a basic measurement that is only available in Spectral Doppler Mode. You can specify a point in
a Spectral Doppler image and trace a curve from that point to calculate the velocity, integral value and
average velocity of blood ow.
1. Press Caliper and select D Trace under soft menu [1]. ‘D Trace is shown in the user information area.
2. Trace a curve. The tracing method is identical to ‘Line Trace.
3. When the measurement is nished, its result is shown on the screen.
X
PSV: Peak Systolic Velocity
X
EDV: End Diastolic Velocity
X
RI: Resistivity Index
X
PI: Pulsatility Index
X
S/D: Ratio of PSV to EDV
X
Vmean: Average velocity.