1-10
Operation Manual
Approved Cables, Transducers and Accessories for EMC
■
Approved Cable for Electromagnetic Compliance
Cables connected to this product may aect its emissions;
Use only the cable types and lengths listed below table.
Cable Type Length
DVI Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
MIC Unshielded Any
Printer Remote Unshielded Any
Audio R.L Shielded Normal
■
Probes
The transducers listed in Chapter 9. ‘Probes’ when used with this product, have been tested to
comply with Group 1 Class B emission as required by International Standard CISPR 11.
■
Approved Accessories for Electromagnetic Compliance
The accessories used with this product may aect its emissions.
CAUTION:
When connecting other customer-supplied accessories to the system, it is the user’s re-
sponsibility to ensure the electromagnetic compatibility of the system. Use only CISPR 11 or CISPR
22, CLASS B compliant devices.
WARNING:
Use of unapproved cables, probes or accessories may increase the electromagnetic
emission or decrease the durability of the ultrasound product.