Samsung MySono U5 Manual de usuario Pagina 165

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 424
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 164
Chapter 5 Measurements and Calculations
5-5
Optimizing Measurement Accuracy
2D Mode
X Resolution is proportional to the frequency of the probe.
X Penetration is inversely proportional to the frequency of the probe.
X Resolution is at its highest when the narrowest ultrasound beam is used at the probe’s focal point.
X Measurements are most accurate at the set focal depth. The further it gets from the focal point,
the less accurate the measurements become, as the beam width increases.
X Using the zoom function or minimizing the depth display makes distance or area measurements more
accurate.
M Mode
X The accuracy of time measurements can be increased when the sweep velocity and the display
format are set to high values.
X The accuracy of distance measurements can be increased when the display format is set to
higher values.
Doppler Mode
X It is recommended to use lower frequency ultrasound for measurement of faster blood ows.
X
The size of the sample volume is limited by the axial direction of the ultrasound.
X Using lower frequency ultrasound increases penetration.
X The accuracy of time measurements can be increased when the sweep velocity is increased.
X The accuracy of velocity measurements can be increased when the vertical scale is set to smaller
values.
X It is most important to use an optimal Doppler angle to enhance the accuracy of velocity
measurements.
Vista de pagina 164
1 2 ... 160 161 162 163 164 165 166 167 168 169 170 ... 423 424

Comentarios a estos manuales

Sin comentarios