Samsung SONOACE R5 Manual de usuario Pagina 27

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 454
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 26
1-11
Chapter 1 Safety
Approved Cables, Probes and Peripherals for EMC
Cables
Cables connected to this product may aect its emissions; Use only the cable types and lengths
listed below table.
Cable Type Length
DVI Shielded Normal
Parallel Shielded Normal
RS232C Shielded Normal
USB Shielded Normal
LAN(RJ45) Twisted pair Any
S-Video Shielded Normal
Foot Switch Shielded 2.5m
B/W Printer Unshielded Coaxial Normal
Printer Remote Unshielded Any
Audio R.L Shielded Normal
VHS Shielded Normal
Probes
The image probe used with this product may aect its emission. The probe listed in ‘Chapter 9.
Probes’ when used with this product, have been tested to comply with the group1 Class A emission
as required by International Standard CISPR 11.
Peripherals
Peripherals used with this product may eect its emissions.
CAUTION: When connecting other customer-supplied accessories to the system, it is the users
responsibility to ensure the electromagnetic compatibility of the system.
WARNING: The use of cables, probes, and peripherals other than those specified may result in
increased emission or decreased Immunity of the Ultrasound System.
Vista de pagina 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 453 454

Comentarios a estos manuales

Sin comentarios