1-10
User Manual
EMC
The testing for EMC(ElectroMagnetic Compatibility) of this system has been performed according to the
international standard for EMC with medical devices (IEC 60601-1-2). This IEC standard was adopted in
Europe as the European norm (EN 60601-1-2).
Guidance and manufacturer’s declaration - electromagnetic emission
This product is intended for use in the electromagnetic environment specied below. The customer or
the user of this product should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment -guidance
RF Emission
CISPR 11
Group 1
The Ultrasound System uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emission
CISPR 11
Class A
The Ultrasound System is suitable for use in all
establishments other than domestic, and may be used in
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes, provided the
following warning is heeded:
Warning: This system is intended for use by healthcare
professionals only. This system may cause radio interference
or may disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as re-
orienting or relocating the Ultrasound System or shielding
the location.
Harmonic Emission
IEC 61000-3-2
Class A
Flicker Emission
IEC 61000-3-3
Complies