7-12
User Manual
Tips!
Setting the Biopsy Guideline
NOTE:
X
If the system is rebooted, the biopsy guideline settings are restored to the
default.
X
Make sure to adjust the biopsy guideline before using the biopsy feature.
X
Support for Multiple Guidelines
If the probe supports multiple biopsy angles, the user may select the biopsy guide angle as
desired.
X
Starting and Finishing a Biopsy
1. Tap Biopsy Guide on the touch screen.
2. Insert the needle along the guideline. And then perform the biopsy as desired.
3. When all procedures are complete, tap Biopsy Guide again to turn this Off. Biopsy will end.
Biopsy Guide Edit Mode
NOTE:
X
Only available in 2D mode and Single mode.
X
Not available in Dual Live or Freeze state.
X
Not available in Write Zoom or Read Zoom state.
X
Starting and Ending Biopsy Guide Edit Mode
1. Tap Biopsy Guide on the touch screen.
2. Use the Angle dial-button on the control panel to enter Biopsy Guide Edit mode.
3. When you enter Biopsy Guide Edit mode, Angle Offset appears.
4. Rotate the Angle dial-button to select Angle Offset or Lateral Offset.
− Angle Offset: This is the angle for the currently selected biopsy guide; the maximum
adjustable offset is the guideline’s outer line angle. Adjustments can be made in 0.1°
increments.
− Lateral Offset: This is the lateral direction for the currently selected biopsy guide.
Adjustments can be made in 0.1 mm increments.
5. Press the Angle dial-button to exit Biopsy Guide Edit mode.