Chapter 5.Measurements and Calculations 5 - 7
Measurement Accuracy Table
The following tables show the accuracy of the measurements available using the product.
Ensure that the results of measurement accuracy checks are kept within the ranges specified in
the table. Except for certain applications or probes, the following accuracy ranges should be
maintained for measurement of a straight distance.
NOTE
To ensure accurate measurements, an accuracy check should be performed at
least once per year. If the measurement accuracy falls outside the ranges
specified in the following table, contact Medison Customer Service.
▐ Modes Performance
Modes
performance
Description
Test
Methodology
Accuracy
(1)
Based on
Specification
Frequency Depth
Penetration Probe
Linear
Phantom Acquisition
<± 2MHz <± 10cm
Phased Array <± 4MHz <± 8cm
Endo-Cavity <± 5.5MHz <± 5cm
Convex <± 7.5MHz <± 3cm
Sensitivity M , PW , CW Phantom Acquisition
(1) Displays the depth with ‘cm’,
(2) Based on center frequency of each probe.
▐ 2D Mode
Measurements:
System Tolerance:
Test
Accuracy
(1)
Range
(2)
Axial Distance: < +/- 4% or 2mm Phantom Acquisition .01 – 25 cm
Lateral Distance: < +/- 4% or 2mm Phantom Acquisition .01 – 35 cm
Axial Resolution: < +/- 4% or 3mm Phantom Acquisition .01 – 25 cm
(1) Measurement accuracy depends on the table and human errors.