5 - 6 SONOACE X6 Operation manual
- It is not recommended to use images in Color/Power Doppler Mode for measurement of
accurate blood flow velocity.
- The amount of blood flow is calculated based on the average velocity rather than the
peak velocity.
- In all applications, the amount of blood flow is measured in PW/CW Spectral Doppler
Mode.
Cursor Position
- All measurements are affected by input data.
- To ensure accurate positioning of the cursor:
Adjust the images on the screen so that they are displayed at maximum granularity.
Use the front edge or boundary point of a probe to make the start and end points of a
measurement object more distinct.
Make sure that the probe direction is always aligned during measurement.