Chapter 1 Safety
Is the interference intermittent or constant?
Does the interference show up only with one transducers operating at the same frequency or with
several transducer?
Do two dierent transducers operating at the same frequency have the same problem?
Is the interference present if the system is moved to a dierent location in the facility?
The answers to these questions will help determine if the problem reside with the system or the scanning
environment. After you answer the question, contact your local Samsung Medison customer service
department.