5-26
Operation Manual
Stenosis Measurement
You can measure the stenosis of each blood vessel system by measuring and calculating an area or
distance.
% StA
Measure the area of the inner and outer walls of a blood vessel. StA stands for Stenosis Area.
1. Select the %StA menu and the rst cursor will appear in 2D Mode.
2. Measure the area of the vessel’s outer wall using the Circ/Area measurement method.
3. When the second cursor appears, measure the area of the vessel’s inner wall under stenosis.
%Stenosis Area. = (Outer Area – Inner Area) / Outer Area × 100
% StD
Measure the diameter of a blood vessel. StD stands for Stenosis Distance.
1. Select the %StD menu and the rst cursor will appear in 2D Mode.
2. Measure the total diameter of a blood vessel using the Distance measurement method.
3. When the second cursor appears, measure the diameter of the vessel’s inner wall under stenosis.
%Stenosis Dist. = (Outer Distance – Inner Distance) / Outer Distance × 100
Heart Rate Measurement
HR (Heart Rate)
You can calculate heart rates for a certain period of time.
1. Select HR in the Measurement menu. A bar appears allowing you to specify a measurement area.
2. Specify the measurement range.
Place the bar at a desired position with the Trackball, and press the Set button.
3. The system measures the heart rate within the specied range automatically. The measurement
results are displayed on the screen.
Vista de pagina 187
1 2 ... 183 184 185 186 187 188 189 190 191 192 193 ... 267 268

Comentarios a estos manuales

Prof Akhigbe 24 Apr 2024 | 14:29:06

I want to learn how to do angle correction on PW Doppler on my machine.

Prof Akhigbe 24 Apr 2024 | 14:30:30

How do I assess PW angle cursor?

Prof Akhigbe 24 Apr 2024 | 14:40:21

I don't want to join the group.