4-7
Chapter 4 Diagnosis Modes
2D Mode Soft Menu
The items that are commonly used in 2D Mode during scanning are provided as Soft Menu items.
You can use the
Soft Menu
dial-button corresponding to a specic menu item on the control panel.
If you rotate the dial-button, the setting for the selected item changes.
Dynamic Range / Harmonic
Use the Soft Menu dial-button [1].
X Dynamic Range
The contrast of an image is adjusted by adjusting the ratio of the minimum and maximum values of
the input signals. Select a value between 50 and 200 by rotating the
Soft Menu
dial-button
[1]
. The
larger the ratio, the smoother the image gets.
X Harmonic
When pressing the
Soft Menu
dial-button
[1]
, the HAR indication appears on the image information.
The OHI (Optimal Harmonic Imaging) function, which allows you to optimize images by using high
frequencies, is provided.
NOTE: The Harmonic function can be used with specific probes.
Reject Level / 2D/C Live
Use
Soft Menu
dial-button
[2]
.
X Reject Level
Noise or echo is removed to make the image clearer. Select a value between 1 and 32 by rotating the
Soft Menu
dial-button
[2]
.
X 2D/C Live
Press the
Soft Menu
dial-button
[2]
. You can observe a scanned area in 2D and Color Doppler images
simultaneously in real time.