8-64
Operation Manual
Dist 20
Traces the cardiac circumference and then draw the cardiac axis. The system automatically draws 20
straight lines perpendicular to the axis and calculates its volume.
NOTE:
X
MR dp/dt can be measured only at 1-3m/s.
X
In Dual 2D Mode, two images can be viewed simultaneously.
X
For RVAWd, RVIDd, RVAWs, and RVIDs, see the LV measurement method.
X
MPA Diam, RPA Diam, and LPA Diam are measured under Aortic Valve Level in Parasternal
Short Axis.
X
C Mode is mainly used for measuring reverse cardiac blood flow.
X
As PISA-Radius or PISA-Alias Vel. measurements require Velocity values, you have to select
color display for Velocity or Vel + Var in C Mode. For more information, refer to the ‘Color Doppler
Mode’ section in Chapter 7. ‘Diagnosis Mode’.
X
Tissue Doppler can be measured in TDI Mode.