Operation Manual
Review the Result of EFW Calculation
1. Press User Key 1 (or User Key 2) on the control panel. The measured items taken for calculating
EFW will be displayed on screen together with the result.
2. To remove the results, press the Clear button on the control panel.
CAUTION:
X
To calculate the GA and the EFW by measuring a specific part of a fetus, an accurate reference
must be selected.
X
If the EFW Reference of the UGEO H60 is changed, the existing EFW is recalculated with the new
reference and then shown on the report. Therefore, it is recommended that you do not change
the EFW Reference of a patient.
X
Since multiple GAs and EFW References are provided, the specialist must make a choice based
on his or her clinical judgment. Choosing an inappropriate reference may result in incorrect
measurements.
X
When performing an exam, one must bear in mind that measurements may vary depending on
the posture of the fetus.
X
In cases where there are multiple fetuses, be sure to verify the Fetus ID to avoid confusion.
X
The GA and EFW Reference possess national and regional characteristics, which must be taken
into consideration when selecting a reference