9-39
Chapter 9 Probes
Disinfection
CAUTION:
Plastic biopsy kits can only be disinfected using a chemically compatible cold-disinfectant.
Disinfection by autoclaving or by using gas or radiation will cause damage to these parts.
Please refer to the user manual of the disinfectant for storage, use, and disposition of the disinfectant.
1. Check the disinfection duration (generally 10 hours) and temperature of the disinfectant.
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method used.
3. Inspect the components for damage such as cracks, rust or breakage. If damage is evident, discontinue
use of the biopsy kit and contact Samsung Medison
Customer Service Department.