5 - 6 SA8000 SE Operation manual
6. A new cursor appears for a new measurement. If more than four distance
measurements are performed, the distance measurements are calculated in order.
The ratio of the first distance to the second, and the ratio of the third to the fourth are
calculated and displayed as percentages.
2D Hip Joint Measurement
This measurement is available in all diagnosis modes.
1. Press the Caliper button, and the measurement item appears on the left side of the
screen.
2. Select ‘2D Hip Joint’ by pressing Flexible Soft button [4] 2D Hip Joint or by
pressing Change or Caliper button several times.
3. Use the Trackball to position the cursor on the start and end points of the first
straight line of Hip Joint. Press the Set button to fix the points.
4. Pressing the Change button before pressing Set button allows the measurement
area to be reset.
5. Set the second and third lines of the Hip Joint in the same way.
6. After position setup, the measurement results are displayed on the screen.
7. A new cursor appears for a new measurement. If more than four distance
measurements are performed, the distance measurements are calculated in order.
Apply the alpha angle (α ) between the first line and the second line, and the beta angle
(β ) between the first and the third to the table shown below, and calculate the type.
TYPE
α β
1a
60 α< 90 0 β< 55
1b
60 α< 90 55 β< 90
2a/b
50 α< 60 0 β< 90
2c
43 α< 50 77 β< 90
d
43 α< 50 0 β< 77
3/4
0 α< 43